Features:
1. Effortless Operation: The machine is designed to operate automatically, eliminating the need for constant supervision or manual intervention.
2. Compact Design: Requires minimal land space, and boasts a reduced height with a maximum Height≤2500mm, ensuring it fits easily into most laboratory settings.
3. Rapid Performance: Achieves quick liquid nitrogen production, taking merely 1.5 hours from start-up to output, ensuring speedy availability for your research needs.
4. Proven Technology: Ensuressafe, reliable, and easy-to-maintain operation with mature technology. The use of ordinary refrigerator compressors enables easy repairs by technicians familiar with refrigerator systems.
With ordinary refrigerator compressors on the market, people who can repair refrigerators will be able to repair it.
5.Maintenance-Free Convenience: Experience continuous operation without the hassle of downtime maintenance or defrosting, maximizing research productivity.
Nitrogen Liquefier: Innovative Working PrincipleEmbracing advanced cryogenic mixed refrigerant technology,
this nitrogen liquefier integrates a state-of-the-art regenerative multi-component mixed refrigerant throttle refrigeration cycle. Carefully chosen high, middle, and low boiling point components form a multi-component refrigerant system. Nitrogen liquefier. The innovative design allows for efficient cooling from ambient temperature down to a crisp -196 °C. The relay matching of effective refrigeration temperature zones of each boiling point component ensures optimal performance.Thesystem achieves the precise purpose of nitrogen liquefaction (-180 °C @ 0.6 MPa) effortlessly.
Features:
1. Seamless Automation: The machine's automatic running capability demands no constant attention, making it highly efficient.
2. Space-Saving: Occupies minimal land space, with a sleek, low-profile design (maximum Height≤2500mm)
3. Swift Liquid Nitrogen Output: Generates liquid nitrogen in just 1.5 hours from start-up, ensuring timely supply for your research processes.
4. Advanced, Reliable Technology: Safeoperation with a reliable system that is easy to maintain. Ordinary refrigerator compressors used in this system allow for straightforward repairs by any refrigerator technician.
With ordinary refrigerator compressors on the market, people who can repair refrigerators will be able to repair it.
Figure 1. Detailed Flow Diagram for the Nitrogen Liquefier from HUILIN
HUILIN Nitrogen Liquefier Specification
Model | WG-STD-05L | WG-STD-1L | WG-STD-3L | WG-STD-5L | WG-STD-10L | WG-STD-20L | WG-STD-25L | WG-STD-30L | WG-STD-40L | WG-STD-50L |
L.Nitrogen Output (L/day) |
12 | 24 | 80 | 120 | 240 | 500 | 360 | 720 | 1000 | 1200 |
L.Nitrogen Output (L/h) |
0.5 | 1 | 3 | 5 | 10 | 20 | 25 | 30 | 40 | 50 |
Power (KW) |
1 | 5 | 8 | 9 | 11 | 25 | 25 | 25 | 35 | 35 |
N2 consumption (Nm3/h) |
≥1 | ≥1.5 | ≥4 | ≥6 | ≥12 | ≥30 | ≥35 | ≥40 | ≥50 | ≥60 |
N2 Purity | 99.9% | 99.9% | 99.9% | 99.9% | 99.9% | 99.9% | 99.9% | 99.9% | 99.9% | 99.9% |
Cooling form | Air | Air | Air | Air | Water | Water | Water | Water | Water | Water |
_Cooling water consumption (Ton/h) |
- | - | - | - | ≥4 | ≥8 | ≥8 | ≥8 | ≥15 | ≥15 |